Load Torque Measurement

Medical connector measurement system “PT-1946MC”

Outline
“PT-1946MC” can perform fitting and torsion stress test specified in common test method of former ISO 594 and ISO 80369 Part 20.
Fitting and torsion stress test is performed by fixing the fitting gauge on the driving unit side and the medical connector on the torque sensor side and twisting while pressing the fitting gauge and the medical connector. Pressing is done by the electro-pneumatic regulator and the pressing load value is measured by the load cell. Torsion is done by the drive unit and the torsion torque is measured with the torque sensor.

PT-1946MC

Medical connector measurement system “PT-1946MC”



Configuration

PT-1946MC_Configuration

Before the test
Before the test
During the test
During the test


Feature
●”PT-1946MC” can perform fitting and torsion stress test specified in common test method of former ISO 594 and ISO 80369 Part 20.
・fitting and torsion stress test specified of ISO80369-3 and ISO8039-6 , ISO80369-7
・ISO80369-20:Common test methods
・former ISO 594

●Panel screen.
A panel screen is provided on the screen of the computer to display torque, load and angle.

ISO80369_Panel screen

●Graph screen.
Torque, load, angle curve is displayed.
Sampling is performed at specified time intervals.

ISO80369_Graph screen

●Save measurement data as CSV file in text format.

●Compact and lightweight.

●Handy handle is attached for easy carrying.





Specification
Torque sensor 1Nm
Torque resolution1mNm
Total torque accuracy±1%±3Digit

Lord cell 100N
Load resolution0.1N
Load Accuracy±1%±3Digit



Option
A reference connector is required to perform tests conforming to ISO 80369 and former ISO 594 using this device.It is possible to prepare an optional jig for connecting this unit to the reference connector.
※No reference connector is attached to this.Please purchase separately.

≪Jig Production Example≫

ISO80369_JIG
Former ISO594-2 Fig.5
ISO80369_JIG
Former ISO594-2 Fig.6

ISO80369_JIG
Former ISO594-2 Fig.7.8



Corresponding test item
Items in red can be tested with this unit.
※Leak test is not supported.

ISO80369
Small-bore connectors for liquids and gases in healthcare applications.
Part20:Common test methods.
TEST METHODTEST METHOD
FITTINGTEST
Annex B
Leakage by pressure
decay TEST METHOD
Axial force:26.5N~27.5N
Torque:0.08Nm~0.12N
Time:5s~6s
(Abridgement)
Annex C
Falling drop positive-pressure
liquid leakage TEST METHOD
(Abridgement)
Annex D
Subatmospheric-pressure air
leakage TEST METHOD
(Abridgement)
Annex E
Stress cracking TEST METHOD
(Abridgement)
Annex F
Resistance to separation from
axial load TEST METHOD
Hold the axial force for the hold
period from the relevant
APLICATION part of ISO80369.
Annex G
Resistance to separation
from unscrewing TEST METHOD
Apply an unscrewing torque
to the collar as specified
in the relevant APPLICATION
part of ISO80369.
Annex H
Resistance to overriding
TEST METHOD
Apply the torque specified
in the relevant
APPLICATION part of
ISO80369 to the CONNECTOR
under test.
Annex I
Disconnection by unscrewing
TEST METHOD
Apply an unscrewing torque
to the CONNECTOR under
test at a rate of
approximately 2 r/min until
the CONNECTION separates.
※Please refer to ISO 80369-20, ISO 80369-3, ISO 80369-6, ISO 80369-7 for details of the test.


Former ISO594
Conical fitting with a 6%(Luer) taper for syringes,needles and certain other medical equipment.
Part1:General requirements

4.Requirements5.TEST METHOD
FITTINGTEST
4.1
Dimensions
5.1.4
Axial force:5N
4.2
Liquid leakage
5.2.1 5.3.1.1 5.5.1
Torque:
Not exceeting 0.1Nm
Rotation:
Not exceeting 90°
Axial force:
27.5N
Time:
5s
5.2.3
(Abridgement)
4.3
Air leakage
5.3.1.2~5.1.7
(Abridgement)
4.4
Separation force
5.4.2
(Abridgement)
4.5
Stress cracking
5.5.2 5.5.3
(Abridgement)
※Please refer to former ISO594 for details of the test.


Former ISO594
Conical fitting with a 6%(Luer) taper for syringes,needles and certain other medical equipment.
Part2:Lock fitting

4.RequirementsTEST METHOD
FITTINGTEST
4.1
Gauging
5.1
(Abridgement)
4.5
Ease of assembly
(Abridgement)
4.2.1
Liquid leakage
5.2.1
Torque:
Not exceeting 0.12Nm
Axial force:
Not exceeting 27.5N
5.2.3
(Abridgement)
4.2.2
Air leakage
5.3.2 5.3.3
(Abridgement)
4.3
Separation force
5.4.2
Axial force:
Not exceeting 35N
Time:
not less than 10s
4.4
Unscrewing torque
5.5.2
Unscrewing torque:0.02Nm
Time: Nor less than 10s
4.6
Resistance to overriding
5.7
Torque:Not less than 0.15Nm
Time:5s
4.7
Stress cracking
5.8.1
Torque:
Not less than 0.12Nm
Axial force:
Not less than 27.5N
Time:5s
5.8.2
(Abridgement)
※Please refer to former ISO594 for details of the test.



Measurement data example
FITTING
ISO80369-20


ISO80369-20_FITTING

ISO80369_FITTING


Resistance to separation from unscrewing TEST METHOD
ISO80369-20 Annex G


ISO80369-20_Resistance to separation from unscrewing TEST METHOD

ISO80369-20_Resistance to separation from unscrewing TEST METHOD


Resistance to overriding TEST METHOD
ISO80369-20 Annex H


ISO80369_Resistance to overriding TEST METHOD

ISO80369_Resistance to overriding TEST METHOD